Infant Radiant Baby Warmer
Home › Products › Baby Warmer › Infant Radiant Baby Warmer
Infant Radiant Baby Warmer
Business Type : Supplier, Retailer
Features:
- Micro Controller based Electronics
- Controls - Servo Skin/ Apgar
- Pre-warm - Use the equipment with 30% Heater Output in Manual Mode.
- Programmable Mute time
- No need of calibration required for temperature probe.
- High Intensity Quartz Heating element for effective Radiation (Ceramic heater Optional)
- Halogen Observation Lamp
- The Heater Module rotates 90° to both sides to facilitate
- X-Ray Procedures.
- The Parabolic Reflector is designed to send uniform heat to the bed surface.
- Memory Backup to restore set temperature and previous mode, in case of power failure.
- Optimum Bed dimensions ensure optimum work space for the nursing staff.
- Comprehensive Alarms and safety cut-offs.
- Alert (Alarm) when sensor is detached from the patient's skin (optional)
- Instant Warming and Uniform heating
Controller
- Electronics : High speed micro controller
- Mode of operation : Servo skin, manual & apgar
- Skin mode control range : 30 Deg. C to 38 Deg. C
- Manual mode timer : 1 min to 15 mins (selectable), 15 mins to 30 mins (optional)
- Apgar timer : Alerts at 1, 5, 10, 15 & 20 mins (selectable)
- Percentage of heater output : 10% to 100% in 10% increments
- Memory back up : Equipment restores the previous settings
- Programmable mute time : User can select the mute time
- Safety heater cut off : If the skin sensor is removed from the baby's skin
- Over temperature heater cut off : If the skin temperature is >39 Deg. C in any skin/manual mode
- Temperature sensor (probe) : Thermistor sensor
- Accuracy of skin sensor : ±0.1 C
- Accuracy of skin probe Interchangeability : 0.2 C
- Calibration of skin sensor (probe) : No need of calibration after Changing new probe
- Heater element : Quartz Infrared heater Ceramic heater (optional)
Other Details
- Power rating : 230V AC, 50Hz, 700 W
- Maximum heater wattage : 600 Watts
- Maximum power requirement : 700 Watts
- Examination light : 12V, 50W Halogen lamp
- X-ray facility : Underneath the bed
- Mattress tilt : ±30 degrees on both sides
- IV pole : Maximum load 2.2 Kg
- Pump mounting pole : Maximum load 2.2 Kg
- Castor : 4” castors 4 Nos. (2 with brake & 2 without brake
- Monitor shelf : 2 Nos. Fixed
- Height adjustable heater module : Available only in nice 2010
- Cabinet : Available only in nice 2010BC
- Operating noise level : <50 dBA
Controller Alarms:
- Check patient If the skin sensor is removed from
- the baby's skin
- Skin temperature high > 1.0 C from the set temperature
- Skin temperature low < 1.0 C from the set temperature
- Skin temperature probe fail Probe short or open
- Over temperature If skin temperature is 39 C
- Heater fail Heater defect
- System fail Electronic component defect
- Power fail AC power interruption
Environmental
- Operating temperature - 15° C to 35° C
- Storage temperature - -20° C to 60° C
- Operating humidity - 5% to 90% RH, non condensing
- Storage humidity - 0% to 99% RH, non condensing
- Altitude - Sea level to 1-9 miles, 3 Km
- Quality - ISO 9001:2000 & ISO 13485:2003
- Electrical & product safety - IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-21
Read More...